Features Options / Accessories Specifications Comparison Table
Quantum Design Oxford - HelioxVT

Extend Your Experiments to 300 mK

HelioxVT

Explore sub-Kelvin physics with HelioxVT, a compact helium-3 sample-in-vacuum insert designed for integration with helium-4 variable temperature inserts (VTIs). Compatible with new or existing TeslatronPT and TeslatronPT Plus, HelioxVT extends their temperature capabilities down to 300 mK.

With a sample temperature range of 300 mK to 300 K, HelioxVT supports research in quantum materials, superconductivity and low-dimensional physics.

Its self-contained design requires no external gas handling system, helping to simplify experimental setup and operation.

KelvinoxJT
Electrical Transport Measurements

High magnetic fields and low temperatures for hall effect and quantum hall effect measurements.

KelvinoxJT
Low Dimensional Physics

IInvestigate the effect of low temperature and high field on low dimensional materials; nanowires, nanotubes, quantum dots and 2d electron gases.

KelvinoxJT
Spintronics

Spin processing and data storage studies using high magnetic fields and low temperatures.

Features

  • Base temperature:  300 mK for 40 hours with no applied heat load
  • Temperature range:  300 mK to 300 K
  • Temperature stability:  ±3 mK below 1.2 K and ±0.1 K above 1.2 K
  • Cool down time:  5 hours when loaded into cold VTI

Why Choose HelioxVT

  • Compatible with 50 mm access VTIs, HelioxVT enables sub-Kelvin measurements on TeslatronPT and TeslatronPT Plus systems.
  • With no liquid helium at the sample plane, HelioxVT is well suited to neutron and X-ray scattering experiments.
  • Designed without a 1 K pot, HelioxVT eliminates the need for additional room-temperature pumps.
  • Operating in a cold-gas environment, HelioxVT requires no liquid helium.
  • Supplied with 24 constantan DC lines as standard, HelioxVT also features an additional line-of-sight port to accommodate experiment-specific wiring requirements.

NEW System Control

Lake Shore Model 346 Cryogenic Temperature Controller

Lake Shore Model 346 Cryogenic Temperature Controller

HelioxVT now benefits from upgraded control electronics and software, combining the Lake Shore Model 346 cryogenic temperature controller with Quantum Design Oxford's cryostat control software, DECS, to deliver a streamlined user experience.

The Lake Shore Model 346 provides precise, stable temperature control, while DECS offers a browser-based interface for convenient system operation, with remote access, continuous data logging, and automated control capabilities.

Together, these provide a modern control platform for HelioxVT, allowing researchers to spend less time on setup and operation, and more time focusing on their research.

Options / Accessories

Specifications

Temperature Range: 300 mK to 300 K
Base Temperature: 300 mK
Hold Time: 40 hours at 300 mK with no applied heat load
Performance Under Load: Below 350 mK for 6 hours with 50 µW applied
Cooldown Time: 5 hours when loaded into a cold VTI
Temperature Stability: ±3 mK below 1.2 K; ±0.1 K above 1.2 K
Sample Environment: Vacuum
Sample Space: 44 mm diameter inner vacuum chamber
VTI Compatibility: 50 mm access

Comparison Table

Probe Temperature Range Cool-Down Time Sample Holder Options Line of Sight Port for Custom Wiring Best For
Basic Probe 1.5 – 300 K 2 hours No Yes Custom
Universal Measurement Probe 1.5 – 300 K 3 hours Yes No Electrical transport/RF/Optical measurements
HelioxVT 0.3 – 300 K 5 hours Yes Yes Lower temperatures